当前位置: 首页 > 有奖问答 >双指数拟合荧光寿命怎么分析

双指数拟合荧光寿命怎么分析

作者 shimaohu123
来源: 小木虫 100 2 举报帖子
+关注

Calculated using 2 exponentials

Prompt data : Prompt
Decay data : Decay

The initial paramters are:

Shift Value = 0 ch
               0 sec
Shift Limit = 40 ch
               8.77915E-09 sec

T1 Estimate = 11.57449 ch
               2.540355E-09 sec
T2 Estimate = 46.29797 ch
               1.016142E-08 sec

A Free
B1 Free
B2 Free

Prompt and decay LO = 404 ch
                       8.866941E-08 sec
Prompt and decay HI = 647 ch
                       1.420027E-07 sec

Background on prompt = 3
Time calibration = 2.194787E-10 sec/ch
---------------------------------------------
The fitted parameters are:

Hi reduced to: 607 ch

SHIFT = 0.8839028 ch
         1.939979E-10 sec
         S.Dev = 3.790324E-12 sec

T1 = 12.29505 ch
      2.698503E-09 sec
      S.Dev = 1.407488E-10 sec
T2 = 18.58244 ch
      4.07845E-09 sec
      S.Dev = 1.984843E-11 sec

A  = 7.525081
      S.Dev = 0.44061

B1 = -0.2825815
      [-72.85 Rel.Ampl]
      S.Dev = 2.110813E-03
B2 = 0.4436326
      [172.85 Rel.Ampl]
      S.Dev = 1.557658E-03

CHISQ = 1.186022
[ 198 degrees of freedom ]

Chi-squared Probability =   3.752265%
Durbin-Watson Parameter =   1.950561
Negative residuals      =   48.52941%
Residuals < 1 s.dev     =   64.21568%
Residuals < 2 s.dev     =   93.13725%
Residuals < 3 s.dev     =    99.5098%
Residuals < 4 s.dev     =    99.5098%



双指数是两个寿命,怎么分析呢,里面的参数怎么看 返回小木虫查看更多

今日热帖
  • 精华评论
  • 102244

    你好,请问这个问题你已经解决了吗?我现在的问题跟上面这个是一样的,我也是不太会分析,请教一下可以帮忙吗?

  • 莘深

    这个可以问仪器的工程师,我问了一下,他说T1和T2都是荧光寿命,你这T1=1.407488E-10 sec,T2=1.016142E-08 sec,都是物质的寿命。顺便问一下楼主,你测的是固体荧光寿命吗?用什么物质做的Prompt,我最近也在侧,用固体样品做了Decay,Prompt没有找到合适的物质做,

猜你喜欢
下载小木虫APP
与700万科研达人随时交流
  • 二维码
  • IOS
  • 安卓