[求助] HPLC,使用出现故障

液相每天开机都有这个页面(plunger homing over pressure),每天清洗完一遍那个过滤器就能用,可是第二天还有这个界面,求助啊。还有我最近用DAD检测器跑出来的标准物质图、对照图和纯溶剂图,几张图都一样(已经确定标准物质的最大吸收波长是280nm),都有我跑的那个标准物质,且响应值差不多,为什么啊?
回复此楼
已阅   回复此楼   关注TA 给TA发消息 送TA红花 TA的回帖

liweixiaoye

至尊木虫 (著名写手)

正股级巡视员

【答案】应助回帖

★ ★ ★ ★ ★
感谢参与,应助指数 +1
lindanana: 金币+5, ★★★★★最佳答案 2012-09-20 09:03:45
“plunger homing over press”。仪器报错的说明应该是:Plunger homing over pressure
AlarmHomingRestriction:
The solvent delivery system detected excessive pressure while moving the plunger forward to its home registration position. This condition may be caused by:
@Failure to remove the pin plug restrictor;
@Clogged in-line filter at the solvent delivery system;
@Blockage in the waste line;
@Excessive restriction between the solvent delivery system and the sample loop;
@Failure of the sample loop to vent the solvent delivery system to atmosphere by opening valves V1, V2 and V3.
    既然是over press,肯定是哪里超过了自检压力,应该是@Excessive restriction between the solvent delivery system and the sample loop;可能性最大。既然是超过了自检压力,泄压应该可以自检通过吧。你可以根据流路,看到泵后的第一个保护柱。
    用手拧开左边进口泄压,再重新自检,应该可以自检通过。最大可能是在线过滤器脏了,拆下来洗洗吧。拆了后打开在线过滤器,把中间的柱芯用甲醇超声清洗干净,再安装上去就可以了(安装有方向)。
人生短短几个秋,不醉不罢休!
2楼2012-09-14 19:48:58
已阅   回复此楼   关注TA 给TA发消息 送TA红花 TA的回帖

zhaozhilei-3208

铜虫 (初入文坛)

【答案】应助回帖

感谢参与,应助指数 +1
说明你的柱子可能被污染了,有残留,每次都冲出来一些,换新柱子试试看
倚天照海花无数,流水高山心自知
3楼2012-09-14 22:04:44
已阅   回复此楼   关注TA 给TA发消息 送TA红花 TA的回帖

xdz1978

至尊木虫 (文坛精英)

持之以恒!

【答案】应助回帖

感谢参与,应助指数 +1
从楼主的情况看,我认为柱子是没问题的,可能是保护柱的问题
一身正气,两袖清风,三餐温饱,四大皆空!
4楼2012-09-14 22:13:58
已阅   回复此楼   关注TA 给TA发消息 送TA红花 TA的回帖

魔术

金虫 (知名作家)

【答案】应助回帖

如果不好查找原因的话,直接电话给仪器公司的工程师,一般会告诉怎么回事,以及如何解决的。
早知道工作后这么难找对象,在学校就该好好恋爱,并且不分手。所谓的谈一场不分手的恋爱是多么美好的事啊!
5楼2012-09-19 14:31:03
已阅   回复此楼   关注TA 给TA发消息 送TA红花 TA的回帖

lindanana

新虫 (著名写手)

引用回帖:
4楼: Originally posted by xdz1978 at 2012-09-14 22:13:58
从楼主的情况看,我认为柱子是没问题的,可能是保护柱的问题

柱子木有问题的,谢谢啦
6楼2012-09-20 09:02:11
已阅   回复此楼   关注TA 给TA发消息 送TA红花 TA的回帖

lindanana

新虫 (著名写手)

引用回帖:
2楼: Originally posted by liweixiaoye at 2012-09-14 19:48:58
“plunger homing over press”。仪器报错的说明应该是:Plunger homing over pressure
AlarmHomingRestriction:
The solvent delivery system detected excessive pressure while moving the plunger forward t ...

谢谢,虫友,我把在线过滤器用37%的稀硝酸洗的,又把仪器用低流速冲了一夜,就好了,thx
7楼2012-09-20 09:05:01
已阅   回复此楼   关注TA 给TA发消息 送TA红花 TA的回帖
相关版块跳转 我要订阅楼主 lindanana 的主题更新
最具人气热帖推荐 [查看全部] 作者 回/看 最后发表
[考研] 材料化工调剂 +10 今夏不夏 2026-03-01 11/550 2026-03-01 19:49 by 无懈可击111
[考研] 306分材料调剂 +4 chuanzhu川烛 2026-03-01 5/250 2026-03-01 19:48 by 无际的草原
[考研] 0856化工专硕求调剂 +12 董boxing 2026-03-01 12/600 2026-03-01 19:45 by 材子momo
[考研] 265分求调剂不调专业和学校有行学上就 +5 礼堂丁真258 2026-02-28 7/350 2026-03-01 19:12 by Js512888
[考研] 298求调剂 +6 axyz3 2026-02-28 6/300 2026-03-01 19:00 by 18137688336
[考研] 272求调剂 +6 材紫有化 2026-02-28 6/300 2026-03-01 18:58 by 18137688336
[考研] 0856材料求调剂 +11 hyf hyf hyf 2026-02-28 12/600 2026-03-01 18:57 by 18137688336
[考研] 291分工科求调剂 +9 science饿饿 2026-03-01 10/500 2026-03-01 18:55 by 18137688336
[考博] 26申博 +4 想申博! 2026-02-26 6/300 2026-03-01 17:32 by 想申博!
[考研] 328求调剂 +3 aaadim 2026-03-01 5/250 2026-03-01 17:29 by njzyff
[考研] 281求调剂 +4 2026计算机_诚心 2026-03-01 7/350 2026-03-01 17:20 by 2026计算机_诚心
[基金申请] 刚录用,没有期刊号,但是在线可看的论文可以放为代表作吗 10+3 arang1 2026-03-01 3/150 2026-03-01 16:43 by babero
[考研] 0856调剂 +4 刘梦微 2026-02-28 4/200 2026-03-01 15:35 by 吸一口猫气
[考研] 课题组接收材料类调剂研究生 +3 gaoxiaoniuma 2026-02-28 4/200 2026-03-01 14:30 by jjj三跨
[考研] 284求调剂 +6 天下熯 2026-02-28 6/300 2026-03-01 14:19 by Ducount.Y
[考研] 311求调剂 +9 南迦720 2026-02-28 10/500 2026-03-01 10:55 by sunny81
[硕博家园] 博士自荐 +6 科研狗111 2026-02-26 10/500 2026-03-01 10:02 by 科研狗111
[论文投稿] Optics letters投稿被拒求助 30+3 luckyry 2026-02-26 4/200 2026-03-01 09:06 by babero
[考研] 272求调剂 +4 田智友 2026-02-28 4/200 2026-03-01 06:43 by 刘兵
[考研] 材料调剂 +4 爱擦汗的可乐冰 2026-02-28 4/200 2026-03-01 00:38 by 猫猫球alter
信息提示
请填处理意见