[求助] HPLC,使用出现故障

液相每天开机都有这个页面(plunger homing over pressure),每天清洗完一遍那个过滤器就能用,可是第二天还有这个界面,求助啊。还有我最近用DAD检测器跑出来的标准物质图、对照图和纯溶剂图,几张图都一样(已经确定标准物质的最大吸收波长是280nm),都有我跑的那个标准物质,且响应值差不多,为什么啊?
回复此楼

» 猜你喜欢

» 本主题相关价值贴推荐,对您同样有帮助:

已阅   回复此楼   关注TA 给TA发消息 送TA红花 TA的回帖

liweixiaoye

至尊木虫 (著名写手)

正股级巡视员

【答案】应助回帖

★ ★ ★ ★ ★
感谢参与,应助指数 +1
lindanana: 金币+5, ★★★★★最佳答案 2012-09-20 09:03:45
“plunger homing over press”。仪器报错的说明应该是:Plunger homing over pressure
AlarmHomingRestriction:
The solvent delivery system detected excessive pressure while moving the plunger forward to its home registration position. This condition may be caused by:
@Failure to remove the pin plug restrictor;
@Clogged in-line filter at the solvent delivery system;
@Blockage in the waste line;
@Excessive restriction between the solvent delivery system and the sample loop;
@Failure of the sample loop to vent the solvent delivery system to atmosphere by opening valves V1, V2 and V3.
    既然是over press,肯定是哪里超过了自检压力,应该是@Excessive restriction between the solvent delivery system and the sample loop;可能性最大。既然是超过了自检压力,泄压应该可以自检通过吧。你可以根据流路,看到泵后的第一个保护柱。
    用手拧开左边进口泄压,再重新自检,应该可以自检通过。最大可能是在线过滤器脏了,拆下来洗洗吧。拆了后打开在线过滤器,把中间的柱芯用甲醇超声清洗干净,再安装上去就可以了(安装有方向)。
人生短短几个秋,不醉不罢休!
2楼2012-09-14 19:48:58
已阅   回复此楼   关注TA 给TA发消息 送TA红花 TA的回帖

zhaozhilei-3208

铜虫 (初入文坛)

【答案】应助回帖

感谢参与,应助指数 +1
说明你的柱子可能被污染了,有残留,每次都冲出来一些,换新柱子试试看
倚天照海花无数,流水高山心自知
3楼2012-09-14 22:04:44
已阅   回复此楼   关注TA 给TA发消息 送TA红花 TA的回帖

xdz1978

至尊木虫 (文坛精英)

持之以恒!

【答案】应助回帖

感谢参与,应助指数 +1
从楼主的情况看,我认为柱子是没问题的,可能是保护柱的问题
一身正气,两袖清风,三餐温饱,四大皆空!
4楼2012-09-14 22:13:58
已阅   回复此楼   关注TA 给TA发消息 送TA红花 TA的回帖

魔术

金虫 (知名作家)

【答案】应助回帖

如果不好查找原因的话,直接电话给仪器公司的工程师,一般会告诉怎么回事,以及如何解决的。
早知道工作后这么难找对象,在学校就该好好恋爱,并且不分手。所谓的谈一场不分手的恋爱是多么美好的事啊!
5楼2012-09-19 14:31:03
已阅   回复此楼   关注TA 给TA发消息 送TA红花 TA的回帖

lindanana

新虫 (著名写手)

引用回帖:
4楼: Originally posted by xdz1978 at 2012-09-14 22:13:58
从楼主的情况看,我认为柱子是没问题的,可能是保护柱的问题

柱子木有问题的,谢谢啦
6楼2012-09-20 09:02:11
已阅   回复此楼   关注TA 给TA发消息 送TA红花 TA的回帖

lindanana

新虫 (著名写手)

引用回帖:
2楼: Originally posted by liweixiaoye at 2012-09-14 19:48:58
“plunger homing over press”。仪器报错的说明应该是:Plunger homing over pressure
AlarmHomingRestriction:
The solvent delivery system detected excessive pressure while moving the plunger forward t ...

谢谢,虫友,我把在线过滤器用37%的稀硝酸洗的,又把仪器用低流速冲了一夜,就好了,thx
7楼2012-09-20 09:05:01
已阅   回复此楼   关注TA 给TA发消息 送TA红花 TA的回帖
相关版块跳转 我要订阅楼主 lindanana 的主题更新
最具人气热帖推荐 [查看全部] 作者 回/看 最后发表
[考研] 291分调剂 +8 上岸小莹加油 2026-04-09 9/450 2026-04-15 00:41 by zzzggc
[考研] 调剂求收留 +33 果然有我 2026-04-10 34/1700 2026-04-15 00:37 by zzzggc
[考研] 071000生物学调剂求助 +18 zzzzwww 2026-04-09 21/1050 2026-04-14 15:39 by zs92450
[考研] 一志愿沪9,326求生物学调剂 +10 刘墨墨 2026-04-13 10/500 2026-04-14 15:16 by zs92450
[考研] 105500药学求调剂 +4 x_skys 2026-04-12 4/200 2026-04-14 13:37 by rndfc
[考研] 材料085601调剂 +32 何润采123 2026-04-10 34/1700 2026-04-14 08:47 by 木木mumu~
[考研] 一志愿2110,化学学硕310分,本科重点双非求调剂 +20 努力奋斗112 2026-04-08 20/1000 2026-04-13 14:24 by 张zhihao
[考研] 材料考研调剂 +29 云木达达 2026-04-11 31/1550 2026-04-13 13:32 by lyh鲁老师
[考研] 0854调剂 +10 长弓傲 2026-04-11 11/550 2026-04-13 10:38 by wp06
[考研] 求调剂 +16 张番茄不炒蛋 2026-04-10 17/850 2026-04-12 13:58 by 熬夜成!
[考研] 材料与化工300求调剂 +39 肖开文 2026-04-09 43/2150 2026-04-12 01:30 by 秋豆菜芽
[考研] 284求调剂 +11 archer.. 2026-04-09 12/600 2026-04-11 20:23 by 蓝云思雨
[考研] 359求调剂 +5 胃痉挛累了 2026-04-11 5/250 2026-04-11 19:55 by lbsjt
[考研] 085410-273求调剂 +6 X1999 2026-04-10 6/300 2026-04-11 10:32 by Delta2012
[考研] 346,工科求调剂 +3 moser233 2026-04-09 3/150 2026-04-11 10:04 by zhq0425
[考研] 085402通信工程调剂,有4项学科竞赛国奖(电赛国二),硕士研究生调剂自荐信。 +5 m永o不v言o弃m 2026-04-09 5/250 2026-04-11 09:33 by zhq0425
[考研] 一志愿华东师范生物学326分,求调剂 +8 刘墨墨 2026-04-09 8/400 2026-04-10 12:00 by pengliang8036
[考研] 296求调剂 +6 汪!?! 2026-04-08 6/300 2026-04-10 11:02 by mattzhming
[考研] 349学科化学045106求调剂,化学类都可以 +8 保好懂懂 2026-04-08 8/400 2026-04-09 14:03 by xulei3024
[考研] 086000生物与医药调剂 +7 awwwwwooooo 2026-04-09 7/350 2026-04-09 13:31 by 北极159263
信息提示
请填处理意见