24小时热门版块排行榜    

Znn3bq.jpeg
汕头大学海洋科学接受调剂
查看: 1531  |  回复: 6

lindanana

新虫 (著名写手)

[求助] HPLC,使用出现故障

液相每天开机都有这个页面(plunger homing over pressure),每天清洗完一遍那个过滤器就能用,可是第二天还有这个界面,求助啊。还有我最近用DAD检测器跑出来的标准物质图、对照图和纯溶剂图,几张图都一样(已经确定标准物质的最大吸收波长是280nm),都有我跑的那个标准物质,且响应值差不多,为什么啊?
回复此楼

» 猜你喜欢

» 本主题相关价值贴推荐,对您同样有帮助:

已阅   回复此楼   关注TA 给TA发消息 送TA红花 TA的回帖

liweixiaoye

至尊木虫 (著名写手)

正股级巡视员

【答案】应助回帖

★ ★ ★ ★ ★
感谢参与,应助指数 +1
lindanana: 金币+5, ★★★★★最佳答案 2012-09-20 09:03:45
“plunger homing over press”。仪器报错的说明应该是:Plunger homing over pressure
AlarmHomingRestriction:
The solvent delivery system detected excessive pressure while moving the plunger forward to its home registration position. This condition may be caused by:
@Failure to remove the pin plug restrictor;
@Clogged in-line filter at the solvent delivery system;
@Blockage in the waste line;
@Excessive restriction between the solvent delivery system and the sample loop;
@Failure of the sample loop to vent the solvent delivery system to atmosphere by opening valves V1, V2 and V3.
    既然是over press,肯定是哪里超过了自检压力,应该是@Excessive restriction between the solvent delivery system and the sample loop;可能性最大。既然是超过了自检压力,泄压应该可以自检通过吧。你可以根据流路,看到泵后的第一个保护柱。
    用手拧开左边进口泄压,再重新自检,应该可以自检通过。最大可能是在线过滤器脏了,拆下来洗洗吧。拆了后打开在线过滤器,把中间的柱芯用甲醇超声清洗干净,再安装上去就可以了(安装有方向)。
人生短短几个秋,不醉不罢休!
2楼2012-09-14 19:48:58
已阅   回复此楼   关注TA 给TA发消息 送TA红花 TA的回帖

zhaozhilei-3208

铜虫 (初入文坛)

【答案】应助回帖

感谢参与,应助指数 +1
说明你的柱子可能被污染了,有残留,每次都冲出来一些,换新柱子试试看
倚天照海花无数,流水高山心自知
3楼2012-09-14 22:04:44
已阅   回复此楼   关注TA 给TA发消息 送TA红花 TA的回帖

xdz1978

至尊木虫 (文坛精英)

持之以恒!

【答案】应助回帖

感谢参与,应助指数 +1
从楼主的情况看,我认为柱子是没问题的,可能是保护柱的问题
一身正气,两袖清风,三餐温饱,四大皆空!
4楼2012-09-14 22:13:58
已阅   回复此楼   关注TA 给TA发消息 送TA红花 TA的回帖

魔术

金虫 (知名作家)

【答案】应助回帖

如果不好查找原因的话,直接电话给仪器公司的工程师,一般会告诉怎么回事,以及如何解决的。
早知道工作后这么难找对象,在学校就该好好恋爱,并且不分手。所谓的谈一场不分手的恋爱是多么美好的事啊!
5楼2012-09-19 14:31:03
已阅   回复此楼   关注TA 给TA发消息 送TA红花 TA的回帖

lindanana

新虫 (著名写手)

引用回帖:
4楼: Originally posted by xdz1978 at 2012-09-14 22:13:58
从楼主的情况看,我认为柱子是没问题的,可能是保护柱的问题

柱子木有问题的,谢谢啦
6楼2012-09-20 09:02:11
已阅   回复此楼   关注TA 给TA发消息 送TA红花 TA的回帖

lindanana

新虫 (著名写手)

引用回帖:
2楼: Originally posted by liweixiaoye at 2012-09-14 19:48:58
“plunger homing over press”。仪器报错的说明应该是:Plunger homing over pressure
AlarmHomingRestriction:
The solvent delivery system detected excessive pressure while moving the plunger forward t ...

谢谢,虫友,我把在线过滤器用37%的稀硝酸洗的,又把仪器用低流速冲了一夜,就好了,thx
7楼2012-09-20 09:05:01
已阅   回复此楼   关注TA 给TA发消息 送TA红花 TA的回帖
相关版块跳转 我要订阅楼主 lindanana 的主题更新
最具人气热帖推荐 [查看全部] 作者 回/看 最后发表
[考研] 367求调剂 +11 hffQAQ 2026-04-09 11/550 2026-04-14 17:48 by lhj2009
[考研] 材料专业344求调剂 +17 hualkop 2026-04-10 22/1100 2026-04-14 16:21 by sxdj2
[考研] 296求调剂 +8 汪!?! 2026-04-09 9/450 2026-04-14 15:15 by 逆水乘风
[教师之家] 转长聘了 +7 简单化xn 2026-04-13 7/350 2026-04-14 14:50 by xindong
[考研] 材料085601调剂 +32 何润采123 2026-04-10 34/1700 2026-04-14 08:47 by 木木mumu~
[考研] 求调剂 +12 璃茉一定上岸 2026-04-10 13/650 2026-04-14 00:08 by Equinoxhua
[考研] 314求调剂 +24 wakeluofu 2026-04-09 25/1250 2026-04-13 08:58 by lhj2009
[考研] 求调剂288 +7 ioodiiij 2026-04-10 9/450 2026-04-13 08:33 by Hayaay
[考研] 一志愿0807 数一英一 313 有没有二轮调剂 +12 emokidd 2026-04-08 13/650 2026-04-13 08:32 by lhj2009
[考研] 0860004 求调剂 309分 +9 Yin DY 2026-04-08 9/450 2026-04-11 22:55 by dongdian1
[考研] 085400 328分 求调剂 +10 喂你一个大橙子 2026-04-09 14/700 2026-04-11 19:53 by lqspecial
[考研] 本人女孩 +7 吼吼, 2026-04-10 9/450 2026-04-11 14:45 by ACS Nano——
[考研] 085410-273求调剂 +6 X1999 2026-04-10 6/300 2026-04-11 10:32 by Delta2012
[考研] 288求调剂 +15 代fish 2026-04-09 16/800 2026-04-11 10:26 by wwj2530616
[考研] 22408 327分求调剂 +4 韵风kon 2026-04-10 4/200 2026-04-11 09:51 by 猪会飞
[考研] 一志愿东北大学控制工程085406数二英二385,求调剂 +8 Ezra_Zhang 2026-04-09 8/400 2026-04-11 09:15 by 猪会飞
[考研] 22408 366分,本科211,一志愿西工大 +4 Rubt 2026-04-09 4/200 2026-04-10 19:51 by chemisry
[考研] 301求调剂 +5 149. 2026-04-10 5/250 2026-04-10 15:45 by 柴小白
[考研] 求调剂材料科学与工程一志愿985初试365分 +5 材化李可 2026-04-08 5/250 2026-04-09 17:00 by Lilly_Li
[考研] 22408 266求调剂 +11 masss11222 2026-04-07 14/700 2026-04-08 11:06 by yulian1987
信息提示
请填处理意见