液相每天开机都有这个页面(plunger homing over pressure),每天清洗完一遍那个过滤器就能用,可是第二天还有这个界面,求助啊。还有我最近用DAD检测器跑出来的标准物质图、对照图和纯溶剂图,几张图都一样(已经确定标准物质的最大吸收波长是280nm),都有我跑的那个标准物质,且响应值差不多,为什么啊?
回复此楼

» 猜你喜欢

» 本主题相关价值贴推荐,对您同样有帮助:

已阅   回复此楼   关注TA 给TA发消息 送TA红花 TA的回帖

liweixiaoye

至尊木虫 (著名写手)

正股级巡视员

【答案】应助回帖

★ ★ ★ ★ ★
感谢参与,应助指数 +1
lindanana: 金币+5, ★★★★★最佳答案 2012-09-20 09:03:45
“plunger homing over press”。仪器报错的说明应该是:Plunger homing over pressure
AlarmHomingRestriction:
The solvent delivery system detected excessive pressure while moving the plunger forward to its home registration position. This condition may be caused by:
@Failure to remove the pin plug restrictor;
@Clogged in-line filter at the solvent delivery system;
@Blockage in the waste line;
@Excessive restriction between the solvent delivery system and the sample loop;
@Failure of the sample loop to vent the solvent delivery system to atmosphere by opening valves V1, V2 and V3.
    既然是over press,肯定是哪里超过了自检压力,应该是@Excessive restriction between the solvent delivery system and the sample loop;可能性最大。既然是超过了自检压力,泄压应该可以自检通过吧。你可以根据流路,看到泵后的第一个保护柱。
    用手拧开左边进口泄压,再重新自检,应该可以自检通过。最大可能是在线过滤器脏了,拆下来洗洗吧。拆了后打开在线过滤器,把中间的柱芯用甲醇超声清洗干净,再安装上去就可以了(安装有方向)。
人生短短几个秋,不醉不罢休!
2楼2012-09-14 19:48:58
已阅   回复此楼   关注TA 给TA发消息 送TA红花 TA的回帖

zhaozhilei-3208

铜虫 (初入文坛)

【答案】应助回帖

感谢参与,应助指数 +1
说明你的柱子可能被污染了,有残留,每次都冲出来一些,换新柱子试试看
倚天照海花无数,流水高山心自知
3楼2012-09-14 22:04:44
已阅   回复此楼   关注TA 给TA发消息 送TA红花 TA的回帖

xdz1978

至尊木虫 (文坛精英)

持之以恒!

【答案】应助回帖

感谢参与,应助指数 +1
从楼主的情况看,我认为柱子是没问题的,可能是保护柱的问题
一身正气,两袖清风,三餐温饱,四大皆空!
4楼2012-09-14 22:13:58
已阅   回复此楼   关注TA 给TA发消息 送TA红花 TA的回帖

魔术

金虫 (知名作家)

【答案】应助回帖

如果不好查找原因的话,直接电话给仪器公司的工程师,一般会告诉怎么回事,以及如何解决的。
早知道工作后这么难找对象,在学校就该好好恋爱,并且不分手。所谓的谈一场不分手的恋爱是多么美好的事啊!
5楼2012-09-19 14:31:03
已阅   回复此楼   关注TA 给TA发消息 送TA红花 TA的回帖

lindanana

新虫 (著名写手)

引用回帖:
4楼: Originally posted by xdz1978 at 2012-09-14 22:13:58
从楼主的情况看,我认为柱子是没问题的,可能是保护柱的问题

柱子木有问题的,谢谢啦
6楼2012-09-20 09:02:11
已阅   回复此楼   关注TA 给TA发消息 送TA红花 TA的回帖

lindanana

新虫 (著名写手)

引用回帖:
2楼: Originally posted by liweixiaoye at 2012-09-14 19:48:58
“plunger homing over press”。仪器报错的说明应该是:Plunger homing over pressure
AlarmHomingRestriction:
The solvent delivery system detected excessive pressure while moving the plunger forward t ...

谢谢,虫友,我把在线过滤器用37%的稀硝酸洗的,又把仪器用低流速冲了一夜,就好了,thx
7楼2012-09-20 09:05:01
已阅   回复此楼   关注TA 给TA发消息 送TA红花 TA的回帖
相关版块跳转 我要订阅楼主 lindanana 的主题更新
最具人气热帖推荐 [查看全部] 作者 回/看 最后发表
[考研] 295分求调剂 +12 ?要上岸? 2026-04-10 12/600 2026-04-12 12:49 by 学zh
[考研] 308求调剂 +5 VvvvL 2026-04-10 5/250 2026-04-12 10:17 by babysonlkd
[考研] 22408调剂315分 +3 zhuangyan123 2026-04-09 3/150 2026-04-12 00:25 by 蓝云思雨
[考研] 求调剂 +11 月@163.com 2026-04-07 13/650 2026-04-11 22:55 by BruceLiu320
[考研] 调剂 +6 青灯不负 2026-04-09 6/300 2026-04-11 20:35 by dongdian1
[考研] 271求调剂 +21 2261744733 2026-04-11 21/1050 2026-04-11 20:19 by Angchemist
[考研] 287求调剂 +15 Fnhc 2026-04-07 21/1050 2026-04-10 19:09 by chemisry
[考研] 293调剂 +25 yj1221 2026-04-08 26/1300 2026-04-10 15:02 by 柴小白
[考研] 08工学 309分求调剂 +6 Yin DY 2026-04-08 6/300 2026-04-10 09:18 by Delta2012
[考研] 085404,285分求调剂 +12 薇薇考研 2026-04-07 14/700 2026-04-09 23:10 by parmtree
[考研] 材料专硕(0856) 339分求调剂 +9 哈哈哈鹅哈哈哈 2026-04-09 10/500 2026-04-09 20:01 by Orcid
[考研] 一志愿西南大学生物学学硕344 求生物学相关调剂/生物与医药 +7 超人不会飞@ 2026-04-08 7/350 2026-04-09 09:35 by gong120082
[考研] 286求调剂 +19 Faune 2026-04-08 20/1000 2026-04-09 08:36 by 哦哦123
[考研] 求调剂 +8 吃口冰激凌 2026-04-07 8/400 2026-04-09 08:03 by 5268321
[考研] 考研求调剂 +4 雯??? 2026-04-08 4/200 2026-04-08 21:44 by 土木硕士招生
[考研] 生物学328分求调剂 +9 闪电kkl 2026-04-08 10/500 2026-04-08 21:42 by liuhuiying09
[考研] 材料考研求调剂总分280 +30 mkjlz1 2026-04-06 35/1750 2026-04-08 21:25 by cyh—315
[考研] 生物医药调剂|SCI中科院三区一作+多项科研成果 +8 likangxing 2026-04-07 11/550 2026-04-08 00:02 by lys0704
[考研] 259求调剂 +5 就爱吃土豆呀呀 2026-04-07 5/250 2026-04-07 22:40 by JourneyLucky
[考研] 308求调剂 +3 终不似从前 2026-04-05 3/150 2026-04-05 22:23 by hemengdong
信息提示
请填处理意见