[求助] 求助!!审稿人提出的关于检测限的问题,不懂是什么意思,求帮助。。。已有3人参与

文章在年前投的BB,现在审稿意见回来了,可是看到这个条意见之后就开始捉急了。看不太懂审稿人的意思,所以无从下手修改。
检测限的计算方法其中一种是LOD=3SB/m,SB代表的是空白值的标准偏差。m为校准曲线的斜率。但是按照这个概念,不知道该怎么回答审稿人提出的问题。
The authors also mention that the lower detection limit was 0.3 pg mL−1 (S/N=3). A magnified calibration plot showing the change in the signal in the pg mL−1
range should definitely be presented anywhere. The authors should have such data because they calculated the detection limit.
Also, it would be very helpful to show response curves for concentrations near the detection limit with a change in the background signal to demonstrate that the concentration of 0.3 pg mL−1 can be measured without any problems.
1.magnified calibration plot 指的是什么曲线呢?
2.背景信号改变时浓度接近检测限的响应曲线这个是什么曲线呢?
不甚感激!!!
回复此楼
只要明天还在,就不悲伤,不气馁!
已阅   回复此楼   关注TA 给TA发消息 送TA红花 TA的回帖

zhenwuhuang

至尊木虫 (文学泰斗)

【答案】应助回帖

★ ★ ★ ★ ★
感谢参与,应助指数 +1
wangqin51919: 金币+5, 有帮助 2014-03-23 20:40:17
The authors should have such data because they calculated the detection limit.
因为他们计算的检出限,作者应该有这样的数据。