核磁信噪比
|
| 核磁信噪比怎么计算?如何更准确的计算? |
» 收录本帖的淘帖专辑推荐
NMR&MRI |
» 猜你喜欢
290调剂生物0860
已经有32人回复
271求调剂
已经有28人回复
105500药学求调剂
已经有4人回复
274求调剂
已经有10人回复
335求调剂
已经有20人回复
化工学硕294分,求导师收留
已经有33人回复
290求调剂
已经有22人回复
291 求调剂
已经有36人回复
食品与营养(0955)271求调剂
已经有16人回复
22408 312求调剂
已经有10人回复
2楼2013-07-20 10:17:31
【答案】应助回帖
|
In theory, signal-to-noise ratio S/N can be done in the following way. Process your spectrum, well phase it and correct the baseline. Locate the strongest peak, integrate over it, (say 0.1 ppm depending the peak width) and the integration is the signal. Now, randomly integrate over N noise regions (i.e., the non-peak regions) of 0.1 ppm wide. The average integration is the noise. signal/noise = S/N Some folks like to figure out S/N using peak height of the strongest peak and the height of the noise regions. They will run into argument of peak-to-peak or root-mean-square noise heights. |