24小时热门版块排行榜    

CyRhmU.jpeg
查看: 1814  |  回复: 24
当前只显示满足指定条件的回帖,点击这里查看本话题的所有回帖

夏天的娃

金虫 (小有名气)

[求助] 气相色谱测含有不挥发液体样品中的挥发物质含量时该如何处理?

大家好!
小虫需要测液体样品的浓度,组分是1个脂肪烃1个芳香烃再加上1个液态盐。液态盐高沸点,300度也不气化,不过我只需要知道2个烃的相对浓度,但是盐流入GC柱子的话肯定会堵柱子呀!!
但是在许多外文文献都有使用GC去测我这种体系的含量,只是在色谱上加了别的东西处理一下,我搜搜了一下相关信息很少呢,好像有用“在线过滤器”,“预柱”,但是供应信息很少,预柱一般是HPLC用的多。因此发帖请教大虫有什么东西能使GC处理我这样的体系呢?另外文献里说的的trap和预柱一般贵么。。5000以上怕是老师不大同意了。。。

附上文献里说的预处理使用的东西:
1.For the samples, the whole of the 不挥发盐 was retained by aSample Textlocated between the injector and the chromatographic column. In this way, the result of the analysis was not affected by the presence of 盐, as we were able to verify experimentally.The trap was periodically cleaned to prevent the IL from coming into the column.
2.The amount of 盐 could not be analyzed via gas chromatography, because of the negligible vapor pressure of 盐; The 盐 was collected in a precolumn (cup linear), to avoid contamination of the analytical column.

谢谢呀谢谢!!

[ Last edited by 夏天的娃 on 2012-8-28 at 18:41 ]
回复此楼
已阅   回复此楼   关注TA 给TA发消息 送TA红花 TA的回帖

大耳毛贼

铜虫 (小有名气)

【答案】应助回帖

感谢参与,应助指数 +1
微量进样还是可以的,再者试试固相微萃取行否。
6楼2012-08-28 23:11:51
已阅   回复此楼   关注TA 给TA发消息 送TA红花 TA的回帖
查看全部 25 个回答

lifucheng

专家顾问 (知名作家)

【答案】应助回帖

感谢参与,应助指数 +1
顶空貌似可以用来分析这种样品吧 嗯 气相色谱柱也有预柱哈 可以问问相关的耗材供应商
2楼2012-08-28 18:46:48
已阅   回复此楼   关注TA 给TA发消息 送TA红花 TA的回帖

夏天的娃

金虫 (小有名气)

引用回帖:
2楼: Originally posted by lifucheng at 2012-08-28 18:46:48
顶空貌似可以用来分析这种样品吧 嗯 气相色谱柱也有预柱哈 可以问问相关的耗材供应商

原来咨询过顶空,国产也要1W~2W多= =!放弃。。
3楼2012-08-28 18:50:45
已阅   回复此楼   关注TA 给TA发消息 送TA红花 TA的回帖

suyin2010

木虫 (小有名气)

【答案】应助回帖

感谢参与,应助指数 +1
那看看能不能先将盐分离出来,用HPLC试试,然后再汽化进入气相色谱中
找到自己要做什么
4楼2012-08-28 20:58:24
已阅   回复此楼   关注TA 给TA发消息 送TA红花 TA的回帖
信息提示
请填处理意见