[求助] 气相色谱测含有不挥发液体样品中的挥发物质含量时该如何处理?

大家好!
小虫需要测液体样品的浓度,组分是1个脂肪烃1个芳香烃再加上1个液态盐。液态盐高沸点,300度也不气化,不过我只需要知道2个烃的相对浓度,但是盐流入GC柱子的话肯定会堵柱子呀!!
但是在许多外文文献都有使用GC去测我这种体系的含量,只是在色谱上加了别的东西处理一下,我搜搜了一下相关信息很少呢,好像有用“在线过滤器”,“预柱”,但是供应信息很少,预柱一般是HPLC用的多。因此发帖请教大虫有什么东西能使GC处理我这样的体系呢?另外文献里说的的trap和预柱一般贵么。。5000以上怕是老师不大同意了。。。

附上文献里说的预处理使用的东西:
1.For the samples, the whole of the 不挥发盐 was retained by aSample Textlocated between the injector and the chromatographic column. In this way, the result of the analysis was not affected by the presence of 盐, as we were able to verify experimentally.The trap was periodically cleaned to prevent the IL from coming into the column.
2.The amount of 盐 could not be analyzed via gas chromatography, because of the negligible vapor pressure of 盐; The 盐 was collected in a precolumn (cup linear), to avoid contamination of the analytical column.

谢谢呀谢谢!!

[ Last edited by 夏天的娃 on 2012-8-28 at 18:41 ]
回复此楼
已阅   回复此楼   关注TA 给TA发消息 送TA红花 TA的回帖

luoyuan121

木虫 (职业作家)

【答案】应助回帖

感谢参与,应助指数 +1
如果不怕麻烦的话,就在衬管里面加玻璃棉,多加一点,然后直接扎针,大概每3天要清洗一次衬管,对色谱柱基本上没影响,我们采用的是填充柱,你要是毛细管柱的话,我没试过
12楼2012-08-29 22:37:38
已阅   回复此楼   关注TA 给TA发消息 送TA红花 TA的回帖

luoyuan121

木虫 (职业作家)

【答案】应助回帖

引用回帖:
16楼: Originally posted by 夏天的娃 at 2012-08-30 09:20:15
我好像太外行了。。好像衬管就在汽化室是吗?样品在衬管里汽化的吧。。。嘿嘿。。。...

衬管在气化室里面,可以用镊子弄出来的,将样品打进去过后,不挥发的固体会吸附在石英棉表面,挥发性的物质会被载气带走,我们实验室所测的样品室碳酸钾中的ACT,用了近三年了,就换过一次柱子,基线什么的基本没变化

» 本帖已获得的红花(最新10朵)

24楼2012-08-30 20:06:33