| 查看: 1417 | 回复: 8 | |||
| 本帖产生 1 个 翻译EPI ,点击这里进行查看 | |||
| 当前只显示满足指定条件的回帖,点击这里查看本话题的所有回帖 | |||
[交流]
求助一段英语翻译
|
|||
|
A previous cLC method, involving large injection volumes (20 μL) and on-column focusing techniques, provided a suitable sensitivity using an Inertsil? C8 capillary column, DAD and gradient elution. This method showed that composition of the focusing solution, mobile phase and pH have remarkable effects on HA separation. In order to improve this separation, an isocratic cLC method compatible with MS detection was developed and the influence of several chromatographic factors on separation quality evaluated. Different reversed-phase packings such as Luna? C8, Luna? C18, Synergi? Max-RP (C12) and Synergi Fusion (C18 and polar embedded group) were tested using UV-DAD detection. These columns were chosen due to their high efficiencies and bonded phase surface coverage. Considering the different particle sizes, flow rate was set at 15 μL min?1 for both Synergi Fusion? and Synergi? Max-RP, 12 and 9 μL min?1 for Luna? C8 and Luna? C18 columns respectively. Chromatographic conditions were optimized using experimental design methodologies such as central composite design. Factors and ranges selected for the optimization of mobile phase composition were ACN (11.6–28.4%), ammonium acetate concentration (13.2–46.8 mM) and buffer pH (3.6–4.4). Response variables were expressed in terms of resolution, calculated at baseline, between the worst resolved peak pair NH–H (Rs,min), and retention time of the last eluting peak (tend). From the obtained equations, it could be concluded that Rs,min was significantly affected by both ACN percentage and pH while tend only by ACN. In addition, the ammonium acetate concentration had no significant effect on separation. Chromatographic separation efficiency was optimized for Rs,min ≥ 2 and minimum tend value. Luna? C8 capillary column provided poor resolution. Luna? C18 and Synergi? Max-RP columns provided similar Rs,min to Synergi? Fusion but higher efficiencies. However, Synergi? Max-RP column provided poor peak symmetry. Therefore, Luna? C18 column was selected for further optimization with the quadrupole MS detector. Taking into account the instrumental limitations of the MS capillary nebulizer, the ammonium acetate concentration in the mobile phase was fixed to 5 mM. Chromatographic separation was optimized using multifactorial design. Factors and ranges selected were ACN (12–18%) and pH (3.6–4.1). For the injection, buffered focusing solutions with 5% MeOH were used. Table 1 includes the values of the experimental responses (Rs,min and tend), which were fitted into the following normalized polynomial equations: (1)Rs,min=1.74?0.51pH?1.35ACN?0.06pH2+0.39pH ACN+0.40ACN2 (2)tend=12.18+0.18pH?11.02ACN+0.76pH2?0.94pH ACN+4.83 ACN2 Determination coefficients were 0.988 and 0.992 for Rs,min and tend respectively, showing the reliability of the equations. In the studied domain, both ACN percentage and pH affected significantly to Rs,min (p values 0.0000 and 0.0002 respectively) (Eq. (1)) while ACN (%) affected only tend (p value 0.0000 and 0.0005 for the single and quadratic term respectively) (Eq. (2)). As can be expected, the interactions between factors (ACN-pH) were significant only for Rs,min response (p = 0.0033). In Fig. 1 it can be observed that the ACN percentage decrease in the mobile phase increases Rs,min and tend. However, when pH decreases, Rs,min increases and tend does not change significantly. The chromatographic separation efficiency was optimized by maximizing Rs,min and minimizing tend values which maximized the desirability function over the selected region. The maximum desirability function yielded a tend = 13.6 min and Rs,min = 2.2 at pH 3.6 and 14.5% ACN. The estimated response surfaces predicted that Rs,min > 2 and tend in the range 12.5–15.0 min could be obtained at pH 3.6–3.7 and 13–15% ACN. As a compromise, pH 3.6 and 13% ACN were selected as optimum values. Under these conditions tend = 15.0 min and Rs,min = 2.3 were expected and experimentally assessed. Finally, ammonium formate 5 mM was also tested and selected to avoid problems into the capillary MS nebulizer. |
» 猜你喜欢
材料与化工(0856)304求B区调剂
已经有6人回复
348求调剂
已经有5人回复
308求调剂
已经有5人回复
299求调剂
已经有6人回复
0856,材料与化工321分求调剂
已经有3人回复
311求调剂
已经有7人回复
085600,材料与化工321分,求调剂
已经有8人回复
一志愿华东理工大学081700,初试分数271
已经有7人回复
085601 材料工程 313分 求调剂
已经有5人回复
22408 359分调剂
已经有3人回复
» 抢金币啦!回帖就可以得到:
西南交通大学医学院招收生物医学工程和材料与化工专业硕士研究生
+1/183
广东唯一石化院校——资源与环境专硕招生
+1/43
国家双一流高校-国家级青年人才课题组博士招生
+2/40
招收2名土木工程方向硕士研究生
+1/38
福建农林大学材料院招收调剂硕士生
+1/38
招收材料、化学类调剂生,专硕学硕均可
+1/33
双一流大学湘潭大学“化工过程模拟与强化”国家地方联合工程研究中心招收各类博士生
+1/17
Alicat 层流压差质量流量控制器在新型硅碳负极流化床的应用- 艾里卡特(Alicat)
+3/13
【2026调剂】南方科技大学-深圳理工大学-中科院 联培086000免疫方向(26.9入学)
+1/11
2026年材料/冶金/化工硕士调剂-上海大学全国重点实验室-激光冶金与硅基材料制备
+1/11
南京医科大学-谭磊课题组-诚招博士后
+1/11
教育部长江学者和创新团队发展计划”入选团队招收 材料,化学与化工博士研究生
+1/5
Top-88悉尼科技大学数据科学/AI 招收2027年入学 校奖 博士生1到2名(国际和本地学生)
+1/5
金属材料先进连接技术
+1/4
中科院科研助理招聘
+1/3
5科大招收材料专业硕士
+1/3
化学工程与技术 国家双一流学科 211高校 国家重点实验室 博士研究生
+1/2
天津大学理学院分离分析及功能材料研究组招收硕士研究生
+1/2
上海交通大学AIMS-Lab招收AI for Science【智能制造】方向2026级博士生
+1/1
天津大学福州校区|学长推荐上岸新思路
+1/1
5楼2010-12-07 16:12:49
3楼2010-12-03 18:43:09
6楼2010-12-07 16:14:54
7楼2010-12-07 17:28:31














回复此楼


