| 查看: 1409 | 回复: 8 | |||
| 本帖产生 1 个 翻译EPI ,点击这里进行查看 | |||
| 当前只显示满足指定条件的回帖,点击这里查看本话题的所有回帖 | |||
[交流]
求助一段英语翻译
|
|||
|
A previous cLC method, involving large injection volumes (20 μL) and on-column focusing techniques, provided a suitable sensitivity using an Inertsil? C8 capillary column, DAD and gradient elution. This method showed that composition of the focusing solution, mobile phase and pH have remarkable effects on HA separation. In order to improve this separation, an isocratic cLC method compatible with MS detection was developed and the influence of several chromatographic factors on separation quality evaluated. Different reversed-phase packings such as Luna? C8, Luna? C18, Synergi? Max-RP (C12) and Synergi Fusion (C18 and polar embedded group) were tested using UV-DAD detection. These columns were chosen due to their high efficiencies and bonded phase surface coverage. Considering the different particle sizes, flow rate was set at 15 μL min?1 for both Synergi Fusion? and Synergi? Max-RP, 12 and 9 μL min?1 for Luna? C8 and Luna? C18 columns respectively. Chromatographic conditions were optimized using experimental design methodologies such as central composite design. Factors and ranges selected for the optimization of mobile phase composition were ACN (11.6–28.4%), ammonium acetate concentration (13.2–46.8 mM) and buffer pH (3.6–4.4). Response variables were expressed in terms of resolution, calculated at baseline, between the worst resolved peak pair NH–H (Rs,min), and retention time of the last eluting peak (tend). From the obtained equations, it could be concluded that Rs,min was significantly affected by both ACN percentage and pH while tend only by ACN. In addition, the ammonium acetate concentration had no significant effect on separation. Chromatographic separation efficiency was optimized for Rs,min ≥ 2 and minimum tend value. Luna? C8 capillary column provided poor resolution. Luna? C18 and Synergi? Max-RP columns provided similar Rs,min to Synergi? Fusion but higher efficiencies. However, Synergi? Max-RP column provided poor peak symmetry. Therefore, Luna? C18 column was selected for further optimization with the quadrupole MS detector. Taking into account the instrumental limitations of the MS capillary nebulizer, the ammonium acetate concentration in the mobile phase was fixed to 5 mM. Chromatographic separation was optimized using multifactorial design. Factors and ranges selected were ACN (12–18%) and pH (3.6–4.1). For the injection, buffered focusing solutions with 5% MeOH were used. Table 1 includes the values of the experimental responses (Rs,min and tend), which were fitted into the following normalized polynomial equations: (1)Rs,min=1.74?0.51pH?1.35ACN?0.06pH2+0.39pH ACN+0.40ACN2 (2)tend=12.18+0.18pH?11.02ACN+0.76pH2?0.94pH ACN+4.83 ACN2 Determination coefficients were 0.988 and 0.992 for Rs,min and tend respectively, showing the reliability of the equations. In the studied domain, both ACN percentage and pH affected significantly to Rs,min (p values 0.0000 and 0.0002 respectively) (Eq. (1)) while ACN (%) affected only tend (p value 0.0000 and 0.0005 for the single and quadratic term respectively) (Eq. (2)). As can be expected, the interactions between factors (ACN-pH) were significant only for Rs,min response (p = 0.0033). In Fig. 1 it can be observed that the ACN percentage decrease in the mobile phase increases Rs,min and tend. However, when pH decreases, Rs,min increases and tend does not change significantly. The chromatographic separation efficiency was optimized by maximizing Rs,min and minimizing tend values which maximized the desirability function over the selected region. The maximum desirability function yielded a tend = 13.6 min and Rs,min = 2.2 at pH 3.6 and 14.5% ACN. The estimated response surfaces predicted that Rs,min > 2 and tend in the range 12.5–15.0 min could be obtained at pH 3.6–3.7 and 13–15% ACN. As a compromise, pH 3.6 and 13% ACN were selected as optimum values. Under these conditions tend = 15.0 min and Rs,min = 2.3 were expected and experimentally assessed. Finally, ammonium formate 5 mM was also tested and selected to avoid problems into the capillary MS nebulizer. |
» 猜你喜欢
一志愿 西北大学 ,070300化学学硕,总分287,双非一本,求调剂。
已经有5人回复
0856求调剂
已经有9人回复
一志愿 南京邮电大学 288分 材料考研 求调剂
已经有3人回复
材料与化工304求B区调剂
已经有5人回复
一志愿北京化工大学材料与化工(085600)296求调剂
已经有5人回复
275求调剂
已经有7人回复
274求调剂
已经有19人回复
总分293求调剂
已经有8人回复
北科281学硕材料求调剂
已经有17人回复
寻找调剂
已经有8人回复
» 抢金币啦!回帖就可以得到:
澳门科技大学创新工程学院诚招博士生
+2/136
山东建筑大学 (市政与环境工程学院)硕士研究生招生啦 静电纺丝纤维与盐水分离应用
+1/90
招收材料与化学相关专业硕士研究生
+1/80
留学--博士招生
+1/51
南方油校——资源与环境专硕招生,化学环境生物类专业均可,就业好,有专业壁垒
+1/48
招收2026级环境科学方向(专业代码083001)硕士研究生1-2名
+1/46
招收材料相关硕士研究生
+2/42
上海大学 “生物有机电子材料及器件”团队硕士、博士研究生招聘
+1/36
临沂大学数学与统计学院招2026研究生
+1/30
澳大利亚昆士兰大学化工学院招收全奖博士生2名
+1/28
南京林业大学-国家级青年人才团队 招2026级博士、硕士(合成化学、植物化学方向)
+1/12
【博士招生】天津理工大学国家杰青王铁课题组招收2026年博士研究生
+1/11
电子信息求调剂
+1/10
化学、化工、化妆品、材料背景考生调剂
+1/8
欢迎调剂报考
+1/8
山东济南省属重点高校生态环境类硕士招生
+1/6
哈尔滨工业大学(深圳)国家级青年人才机械工程、热能与能源等专业26年入学博士招生
+1/5
湖南农业大学 环境科学与工程0830+资源与环境专硕0857 调剂名额
+1/5
天津工业大学双一流高校招博士、硕士、博士后
+1/3
南医大基础医学院招收调剂生(2生信1生物))
+1/3
5楼2010-12-07 16:12:49
3楼2010-12-03 18:43:09
6楼2010-12-07 16:14:54
7楼2010-12-07 17:28:31













回复此楼


