24小时热门版块排行榜    

查看: 650  |  回复: 2

lifuchengol

[交流] 【求助】求助:关于仪器 检测限相关问题? 已有2人参与

请问一个关于检测限(LOD)和定量限(LOQ)的问题
我的一篇文章投到analytical chimica acta之后,编辑一直问一个关于LOD和 LOQ的问题。我定义了检测限为3倍信噪比,定量限为10倍信噪比。我做的时候是标准样品倍比稀释,浓度分别为40, 20, 10, 5, 2.5μg/kg,然后用HPLC检测,得到的LOD5μg/kg,为LOQ10μg/kg。但是编辑认为我既然定义了检测限为3被信噪比,定量限为10 被倍信噪比,那么定量限应该为检测限的3.3倍,可是我的浓度本来就是倍比稀释的啊,信噪比大于10的最低浓度为10μg/kg,信噪比大于3的最低浓度为5μg/kg。
我不知道怎么解释他们才能听懂,还是我应该改变检测限和定量限的定义呢?着急,谢谢好心人!
如下是编辑的comments
In your response to my comments you have written "We defined LOD as the lowest drug concentration in the samples can be measured with a signal to the background signal (S/N) ratio of 3/1(S/N), and LOD as the lowest drug concentration in food samples can be measured with a signal to the background signal (S/N) ratio of 10/1(S/N). After the detection, we found that the LOD was 5 g/kg and the LOQ was 10 g/kg."  If 5 µg/kg gives a signal ise ratio of 3/1, 10 µg/kg cannot give a signal ise ratio of 10/1 (it would give just a signal ise ratio of 6/1) !!! Either your value for the LOD or your value for the LOQ is wrong.
回复此楼
已阅   回复此楼   关注TA 给TA发消息 送TA红花 TA的回帖

pavay998

金虫 (小有名气)


小木虫(金币+0.5):给个红包,谢谢回帖交流
从定义上看,这两个值要先算出噪声值,再用倍率计算。
天若有情天亦老
2楼2010-08-19 08:52:44
已阅   回复此楼   关注TA 给TA发消息 送TA红花 TA的回帖

lim寒竹

金虫 (正式写手)


小木虫(金币+0.5):给个红包,谢谢回帖交流
确实棘手,要不你用信噪比法,不用这种方法(直观法)。
社会犹如一条船,每个人都要有掌舵的准备
3楼2010-08-19 10:49:05
已阅   回复此楼   关注TA 给TA发消息 送TA红花 TA的回帖
相关版块跳转 我要订阅楼主 lifuchengol 的主题更新
普通表情 高级回复 (可上传附件)
信息提示
请填处理意见