24小时热门版块排行榜    

查看: 1297  |  回复: 2
当前只显示满足指定条件的回帖,点击这里查看本话题的所有回帖

swimvc

银虫 (著名写手)

[交流] 【求助】求教,什么情况下会导致做CV图时,扫速快时出峰,扫速慢时不出峰 已有2人参与

如题
今天做修饰电极出现了这样的情况,谁知道什么原因呢?
回复此楼
已阅   回复此楼   关注TA 给TA发消息 送TA红花 TA的回帖

Daaaan

新虫 (初入文坛)


小木虫: 金币+0.5, 给个红包,谢谢回帖
楼主知道什么原因了吗?求教。钠离子电池CV,也发现了这种情况。
3楼2019-09-10 18:54:13
已阅   回复此楼   关注TA 给TA发消息 送TA红花 TA的回帖
查看全部 3 个回答

jiehe2009

金虫 (初入文坛)

is your chemical in the electrode quite easy to desorb?


小木虫(金币+0.5):给个红包,谢谢回帖交流
Don't forget the electrochemical reaction occurs at surface. If the chemical for some reason can't form sort of stong enough interaction with electrode, then CV peak can't be found.  Specifically, if the CV scan speed is high, more reactant is used kinetically, the anodic or cathodic current peak is easy to find. Conversely, the CV scan speed is low, relative low content of reactant at the electrode surface is reached, it is more difficult to find CV peak.
On the other hand, is your electrolyte conductive enough? If not, the electrochemical reaction is going to be more sluglish.