24小时热门版块排行榜    

查看: 1625  |  回复: 6
当前只显示满足指定条件的回帖,点击这里查看本话题的所有回帖

deep_bluesea

木虫 (正式写手)

[交流] 【求助】弱弱的问什么是A类错误已有6人参与

刚开始做晶体,在论坛上看到好多关于A类错误或B类错误,请问什么是A类错误什么是B类错误啊?这些错误是由于什么原因导致的呢?
回复此楼
已阅   回复此楼   关注TA 给TA发消息 送TA红花 TA的回帖

qustxrf

木虫 (小有名气)

★ ★
帝夫(金币+2):谢谢回帖交流 2010-06-08 09:33:38
deep_bluesea(金币+1): 2010-06-08 12:04:48
Maximum theta of reflections used to measure unit cell.
            The following tests will not be performed.
            CELLT_01
DIFF019_ALERT_1_A  _diffrn_standards_number is missing
            Number of standards used in measurement.
DIFF020_ALERT_1_A  _diffrn_standards_interval_count and
            _diffrn_standards_interval_time are missing. Number of measurements
            between standards or time (min) between standards.
DIFF022_ALERT_1_A  _diffrn_standards_decay_% is missing
            Percentage decrease in standards intensity.

这些A类错误是什么意思?怎么解决?


作者:monoclinic

CELL003_ALERT_1_A  _cell_measurement_reflns_used is missing
            Number of reflections used to measure unit cell.
晶胞探测使用的衍射点数
CELL004_ALERT_1_A  _cell_measurement_theta_min is missing
            Minimum theta of reflections used to measure unit cell.
            The following tests will not be performed.
            CELLT_01
晶胞探测的最小theta角
CELL005_ALERT_1_A  _cell_measurement_theta_max is missing
            Maximum theta of reflections used to measure unit cell.
            The following tests will not be performed.
            CELLT_01
晶胞探测的最大theta角

这三个数据应该在给你的信息文件里,如:

_cell_measurement_reflns_used     2919  
_cell_measurement_theta_min       2.14
_cell_measurement_theta_max      28.24




DIFF019_ALERT_1_A  _diffrn_standards_number is missing
            Number of standards used in measurement.
DIFF020_ALERT_1_A  _diffrn_standards_interval_count and
            _diffrn_standards_interval_time are missing. Number of measurements
            between standards or time (min) between standards.
DIFF022_ALERT_1_A  _diffrn_standards_decay_% is missing
            Percentage decrease in standards intensity.
这三个是监测晶体的指标,多数仪器已经不用了。如果你采集数据是用CCD衍射仪,那这三个后面都填 “ ? ” 就可以了,如:

_diffrn_standards_number           ?
_diffrn_standards_interval_count  ?
_diffrn_standards_interval_time    ?
_diffrn_standards_decay_%         ?

[ Last edited by monoclinic on 2009-9-12 at 23:16 ]

作者:244073931

很容易去除这些A类错误,填上这些数据(前三个数据在LS文件的结尾部分2theta最大和最小),后面的几个问题如楼上所说,填上问号或0(有的是0),重新检测,即OK

作者:王一755

很容易去除这些A类错误,填上这些数据(前三个数据在LS文件的结尾部分2theta最大和最小),后面的几个问题如楼上所说,填上问号或0(有的是0),重新检测,即OK  在.om.ls文件最后一行

作者:l_viking

这个。。。知道怎么改的但是不知道怎么说