[交流]
【求助】请教GC-MS溶剂延迟问题
|
||||
|
Agilent GC-MS(6890N /5973N)不分流进样,溶剂延迟3min。是否意味着延迟分流阀开启时间为3min?如果升温程序欲从50度以10度/min的速度升至300度,应如何设置升温程序?是否先保持柱温50度,保持时间要多长? 谢谢! |
» 收录本帖的淘帖专辑推荐
色谱 |
» 猜你喜欢
2025年遐想
已经有4人回复
投稿Elsevier的杂志(返修),总是在选择OA和subscription界面被踢皮球
已经有8人回复
自然科学基金委宣布启动申请书“瘦身提质”行动
已经有4人回复
求个博导看看
已经有18人回复
|
谢谢,学习了。 这问题是 J. Chromatogr. A审稿人问的。看了楼上几位朋友的回复,明白了不少。麻烦几位帮我看看,如何回答? 我的GC升温程序:The GC temperature was programmed as follows: from 50 to 120 ◦C at a rate of 10 ◦C min−1 and then to 300 ◦C at a rate of 3 ◦C min−1, where it was held for 2 min. 审稿人的问题:In splitess mode, does "solvent delay" indicate delay in split valve opening? In such a case, usually oven temperature is maintained at isotherm until valve opens, in the operative conditions reported at lines 103-105, after 3 minutes, temperature has already reached 80°C. |
6楼2009-10-09 15:23:33







回复此楼