24小时热门版块排行榜    

查看: 490  |  回复: 3
当前主题已经存档。
当前只显示满足指定条件的回帖,点击这里查看本话题的所有回帖

wxmyy_chem

木虫 (著名写手)

[交流] 求教CIF两个B类问题的解决方法

准备投ACTA的数据有两个B类错误
PLAT222_ALERT_3_B Large Non-Solvent  H  Ueq(max)/Ueq(min)  4.26 Ratio
PLAT242_ALERT_2_B Check Low  Ueq as Compared to Neighbors for  C11  
________________________________________
PLAT222 Type_3 Test Ueq(max)/Ueq(Min) range for H atoms in non-solvent
________________________________________

This test reports on a larger than usual range of U(eq) values for hydrogen
atoms in the non-solvent/anion part of the structure.
Possibe causes are:
1 - disorder, e.g. in t-butyl moieties.
2 - poor data, not adequate for the refinement of individual displacement
     parameters.
3 - Misplaced hydrogen atoms (i.e. there is no density at the position
     where one of the H-atoms is positioned).

________________________________________
PLAT242 Type_2 Test for Unusually Low U(eq) as compared with bonded neighbors
________________________________________
The U(eq) value of an atom is compared with the average U(eq) for
non-hydrogen atoms bonded to it. Large differences may indicate that the
wrong atom type was assigned (e.g. N instead of O). False alarms may
occur for terminal groups such as the t-butyl moiety.
哪位达人可以帮助一下,留下邮箱我把数据发给达人。谢过
回复此楼
稳稳当当做事,坦坦荡荡做人
已阅   回复此楼   关注TA 给TA发消息 送TA红花 TA的回帖
相关版块跳转 我要订阅楼主 wxmyy_chem 的主题更新
普通表情 高级回复(可上传附件)
信息提示
请填处理意见