[求助] 您好,我想咨询下,气相方法学验证以及液相方法学验证线性与范围... 已有4人参与

您好,我想咨询下,气相方法学验证以及液相方法学验证线性与范围,范围一般在多少

发自小木虫Android客户端
回复此楼

» 猜你喜欢

» 本主题相关价值贴推荐,对您同样有帮助:

已阅   回复此楼   关注TA 给TA发消息 送TA红花 TA的回帖

木易十日十月

金虫 (著名写手)

【答案】应助回帖

感谢参与,应助指数 +1
国家标准有跟检测器有关,紫外一般大于10的四次方,

[ 发自手机版 http://muchong.com/3g ]
4楼2016-01-03 12:56:52
已阅   回复此楼   关注TA 给TA发消息 送TA红花 TA的回帖
查看全部 15 个回答

zongyuan

捐助贵宾 (知名作家)

【答案】应助回帖

感谢参与,应助指数 +1
Determination of Linearity and Range—Linearity should be established across the range of the analytical procedure. It
should be established initially by visual examination of a plot of signals as a function of analyte concentration of content. If
there appears to be a linear relationship, test results should be established by appropriate statistical methods (e.g., by calculation
of a regression line by the method of least squares). Data from the regression line itself may be helpful to provide mathematical
estimates of the degree of linearity. The correlation coefficient, y-intercept, slope of the regression line, and residual
sum of squares should be submitted.
The range of the procedure is validated by verifying that the analytical procedure provides acceptable precision, accuracy,
and linearity when applied to samples containing analyte at the extremes of the range as well as within the range.
ICH recommends that, for the establishment of linearity, a minimum of five concentrations normally be used. It is also recommended
that the following minimum specified ranges should be considered:
Assay of a Drug Substance (or a finished product): from 80% to 120% of the test concentration.
Determination of an Impurity: from 50% to 120% of the acceptance criterion.
For Content Uniformity: a minimum of 70% to 130% of the test concentration, unless a wider or more appropriate range
based on the nature of the dosage form (e.g., metered-dose inhalers) is justified.
For Dissolution Testing: ±20% over the specified range (e.g., if the acceptance criteria for a controlled-release product cover
a region from 30%, after 1 hour, and up to 90%, after 24 hours, the validated range would be 10% to 110% of the label
claim).
The traditional definition of linearity, i.e., the establishment of a linear or mathematical relationship between sample concentration
and response, is not applicable to particle size analysis. For particle size analysis, a concentration range is defined (instrument-
and particle size-dependent) such that the measured particle size distribution is not affected by changes in concentration
within the defined concentration range. Concentrations below the defined concentration range may introduce an error
due to poor signal-to-noise ratio, and concentrations exceeding the defined concentration range may introduce an error due
to multiple scattering.
2楼2016-01-03 12:54:10
已阅   回复此楼   关注TA 给TA发消息 送TA红花 TA的回帖

zongyuan

捐助贵宾 (知名作家)

【答案】应助回帖

3. RANGE
The specified range is normally derived from linearity studies and depends on the intended application of the procedure. It is established by confirming that the analytical procedure provides an acceptable degree of linearity, accuracy and precision when applied to samples containing amounts of analyte within or at the extremes of the specified range of the analytical procedure.
The following minimum specified ranges should be considered:
- for the assay of a drug substance or a finished (drug) product: normally from 80 to 120 percent of the test concentration;
- for content uniformity, covering a minimum of 70 to 130 percent of the test concentration, unless a wider more appropriate range, based on the nature of the dosage form (e.g., metered dose inhalers), is justified;
- for dissolution testing: +/-20 % over the specified range;
e.g., if the specifications for a controlled released product cover a region from 20%, after 1 hour, up to 90%, after 24 hours, the validated range would be 0-110% of the label claim
for the determination of an impurity: from the reporting level of an impurity1 to 120% of the specification;
- for impurities known to be unusually potent or to produce toxic or unexpected pharmacological effects, the detection/quantitation limit should be commensurate with the level at which the impurities must be controlled;
Note: for validation of impurity test procedures carried out during development, it may be necessary to consider the range around a suggested (probable) limit.
- if assay and purity are performed together as one test and only a 100% standard is used, linearity should cover the range from the reporting level of the impurities1 to 120% of the assay specification.
5楼2016-01-03 12:58:30
已阅   回复此楼   关注TA 给TA发消息 送TA红花 TA的回帖

zongyuan

捐助贵宾 (知名作家)

国内这个定义很混乱
7楼2016-01-03 12:59:19