| 查看: 2186 | 回复: 17 | ||||
| 当前只显示满足指定条件的回帖,点击这里查看本话题的所有回帖 | ||||
fzxiaofeixia铁杆木虫 (著名写手)
|
[求助]
有几条审稿意见不太理解,寻求 分析化学 相关高人指点,谢谢 已有2人参与
|
|||
|
1.Calibration Curves: a) The concentration values of the calibration solutions, or alternatively, the calibration scedasticity must be reported. b) It doesn't seem that replicates of the calibration solutions have been produced and injected. In fact, table 3 does not report the measurement error, which is usually calculated by the standard deviation of the calibration curve. This point is particularly important because the analysis of trace measurement error may be greater than the measure itself.对这条两条意见不明白,请高人指点! 2.The authors reported that the intra-day and inter-day precisions were 7.39 and 9.77 % respectively. What are the respectively SD? Specify the abbreviation of RSD... 审稿人想表达的确切意见是? 3.如果发现的审稿人的意见明显是错误的,怎么回复啊?最好有英文模板 ![]() |
» 猜你喜欢
93BebMhtakh前后11位开头都是大写
已经有5人回复
欢迎发来filecode的Mz6后的代码验证其规律
已经有86人回复
感觉是下周放榜了
已经有11人回复
filecode=后面第一个是大写字母
已经有10人回复
时间戳变了,能看出什么问题?
已经有4人回复
哪位老哥知道今年的国自然具体哪一天放榜?
已经有15人回复
今天系统多次维护,明天很可能放榜!
已经有9人回复
时间戳又变了8-15
已经有25人回复
2027广东省杰青
已经有6人回复
快农历七夕节了,轻松一下,男人悄悄话,女施主请不要进来。
已经有3人回复
【答案】应助回帖
★ ★ ★ ★ ★ ★ ★ ★ ★ ★ ★ ★ ★ ★ ★
fzxiaofeixia: 金币+15, ★★★很有帮助 2014-07-08 16:59:25
fzxiaofeixia: 金币+15, ★★★很有帮助 2014-07-08 16:59:25
|
标准曲线:a 标准溶液的浓度值或者选择性的方差应该给出 b 没有确切的证据证明标准溶液是否重复测定。实际上表三没有给出测量误差,可以通过标准曲线的标准偏差进行计算。这一点十分重要,因为痕量分析的测量误差往往高于方法误差。 我觉得审稿人是让你证明你做标准曲线的时候溶液不是用的一份,是每一个浓度进行平行配制,并进行测量。至于b则需要补算一个标准偏差。 日内跟日间精确度是7.39跟9.77.他们各自的偏差是多少,只需要算一下就行 |
13楼2014-07-08 16:13:24
fzxiaofeixia
铁杆木虫 (著名写手)
- 应助: 49 (小学生)
- 金币: 8846
- 红花: 12
- 帖子: 1519
- 在线: 367.3小时
- 虫号: 303688
- 注册: 2006-12-03
- 性别: GG
- 专业: 质谱分析
2楼2014-07-05 12:55:27
brucefan
专家顾问 (著名写手)
-

专家经验: +218 - 应助: 422 (硕士)
- 金币: 2294.8
- 散金: 1720
- 红花: 88
- 帖子: 1423
- 在线: 670.2小时
- 虫号: 3048965
- 注册: 2014-03-14
- 性别: GG
- 专业: 凝聚态物性 II :电子结构
- 管辖: 分子模拟
【答案】应助回帖
★ ★ ★ ★ ★ ★ ★ ★ ★ ★ ★ ★ ★ ★ ★
感谢参与,应助指数 +1
fzxiaofeixia: 金币+15, ★★★很有帮助, 谢谢帮助 2014-07-05 19:51:58
感谢参与,应助指数 +1
fzxiaofeixia: 金币+15, ★★★很有帮助, 谢谢帮助 2014-07-05 19:51:58
|
我搞凝聚态物理的,可能理解得不对,但愿意说出来,供你参考。 1.Calibration Curves: a) The concentration values of the calibration solutions, or alternatively, the calibration scedasticity must be reported. b) It doesn't seem that replicates of the calibration solutions have been produced and injected. In fact, table 3 does not report the measurement error, which is usually calculated by the standard deviation of the calibration curve. This point is particularly important because the analysis of trace measurement error may be greater than the measure itself.对这条两条意见不明白,请高人指点! 第一点是说你应该报道所有定标液的浓度值,或者它们的统计分布(即平均值和标准差)。 第二点是审稿人猜测你实际上只用了一个样品,因为表3中没有报道测量误差(即标准差)。 我的建议,如果你只测了一个样品,要么继续补样品,要么回答审稿人你只测了一个以及为什么认为一个就足够了。如果你测了多个样品,那么也告诉审稿人你测了多个,但没有报道定标液浓度的统计分布以及结果的误差。 2.The authors reported that the intra-day and inter-day precisions were 7.39 and 9.77 % respectively. What are the respectively SD? Specify the abbreviation of RSD... 审稿人想表达的确切意见是让你除了报道两种情况的precision之外还要报道它们的SD(我猜是standard deviation?)。另外,请告诉读者RSD代表什么(我估计是Relative standard deviation)。【注】:我的理解是:你报道的precision其实就是RSD,是个百分比(相对值),但审稿人想知道绝对标准差是多少。 3.如果发现的审稿人的意见明显是错误的,怎么回复啊?最好有英文模板 反驳就是了。例如:We thank the Referee for pointing out this, but we do not agree with her/him regarding this point. ... (Then you can "teach" the Referee something here.) |
3楼2014-07-05 17:38:10
fzxiaofeixia
铁杆木虫 (著名写手)
- 应助: 49 (小学生)
- 金币: 8846
- 红花: 12
- 帖子: 1519
- 在线: 367.3小时
- 虫号: 303688
- 注册: 2006-12-03
- 性别: GG
- 专业: 质谱分析
4楼2014-07-05 19:51:19











回复此楼