fan_xmz

金虫 (初入文坛)

[求助] xrd中的问题求助。。。。求大神解释

X-ray diffraction studies using CuK radiation with Ni filter confirmed the single-phase spinel structure for all the samples.
上面是一篇文献给的这个地方是不是给错了(with Ni filter ),应该把filter改成ferrite?
另外一个问题是测xrd有步长是0.02和0.05的有什么不同,越详细越好,谢谢!
回复此楼

» 猜你喜欢

» 本主题相关价值贴推荐,对您同样有帮助:

喜欢瓜子脸瘦的女生,能白点更好!嘿嘿!
已阅   回复此楼   关注TA 给TA发消息 送TA红花 TA的回帖

lee988325

新虫 (小有名气)

【答案】应助回帖

★ ★ ★ ★ ★ ★ ★ ★ ★ ★ ★ ★ ★ ★ ★
感谢参与,应助指数 +1
kunpeng1782: 金币+10, MN-EPI+1, 谢谢您的回答 2013-07-28 15:52:41
fan_xmz: 金币+5, ★★★★★最佳答案 2013-07-28 19:01:17
The K-absorption-edge (K-to-continuum threshold) for nickel is at 1.488 Angstrom. As a result, nickel absorbs Cu Kb radiation much more strongly than it absorbs Cu Ka radiation. A Ni foil filter is often used to attenuate Cu Kb radiation, and is included in our tube housing.  You will still see double peaks due to Ka1 and Ka2   at least at large scattering angles.

http://xrd-cmmp.blogspot.tw/2011 ... ffect-of-steps.html
when running continuous scans, the step size needs to be chosen according to the peak FWHM for effective peak position determination-- but keep in mind that step size impacts net intensity.
5楼2013-07-28 15:07:54