[求助] 接触角测试 仪器

各位虫友,大家好!
求助一个关于生物膜材料接触角测试方面的问题。
我文中写的方法是Contact angles of the nanocoated film surface were measured by a video contact angle meter (12907666, Kruss, Germany) using the sessile drop method.
现在审稿人关于接触角提出了很多问题,有些问题不是很明白,请教大家怎么回答。
1,ref given for the kruss equipement is the serial number, not the model type. Did you used a DSA serie model, a TVA or a Mobiledrop ???
2,What was the rate of evaporation of the dropplet during measurment time ?? 3,Measurment of contact angle on hydrophilic biopolymer require a lot of consideration, because value may have no signification when absorption occurs during speading or wetting phenonemon.
4,What was the shape of kinetics of angle and volume ? Are there linear or not ?
5,Was kinetic of wetting similar between all samples and between each layer?
回复此楼
已阅   回复此楼   关注TA 给TA发消息 送TA红花 TA的回帖

hwweven

至尊木虫 (知名作家)

医疗器械注册和审核

优秀版主

【答案】应助回帖

★ ★ ★ ★ ★ ★ ★ ★ ★ ★ ★ ★ ★ ★ ★ ★ ★ ★ ★ ★ ★ ★
感谢参与,应助指数 +1
fyjqt: 金币+2, 应助奖励~ 2012-09-13 18:53:46
yuyanggood: 金币+20, ★★★很有帮助 2012-09-15 22:07:53
1、说你给的是仪器的序列号,而不是型号。
2、问你测试时水滴的表面蒸发速率,做接触角是要考虑表面蒸发的,水滴会随测试时间延长明显变小,接触角会逐渐降低,要说明测试所用的时间是否一致。
3、说该亲水表面的接触角数值并不可靠,因为表面会发生渗透吸收,lz最好做一个动态接触角说明该表面的吸水程度。
4和5,不知道lz具体做什么,不太清楚了。
2楼2012-09-13 16:53:53
已阅   回复此楼   关注TA 给TA发消息 送TA红花 TA的回帖

yuyanggood

金虫 (小有名气)

引用回帖:
2楼: Originally posted by hwweven at 2012-09-13 16:53:53
1、说你给的是仪器的序列号,而不是型号。
2、问你测试时水滴的表面蒸发速率,做接触角是要考虑表面蒸发的,水滴会随测试时间延长明显变小,接触角会逐渐降低,要说明测试所用的时间是否一致。
3、说该亲水表面的 ...

我测试时是用的静态接触角。是不是审稿人问的那些都是关于动态接触角仪器的测试要求?比如说蒸发速率,测试的压力、温度、相对湿度等。
3楼2012-09-13 18:29:59
已阅   回复此楼   关注TA 给TA发消息 送TA红花 TA的回帖

hwweven

至尊木虫 (知名作家)

医疗器械注册和审核

优秀版主

【答案】应助回帖

引用回帖:
3楼: Originally posted by yuyanggood at 2012-09-13 18:29:59
我测试时是用的静态接触角。是不是审稿人问的那些都是关于动态接触角仪器的测试要求?比如说蒸发速率,测试的压力、温度、相对湿度等。...

估计审稿人就是想让你补充动态接触角的实验。