24小时热门版块排行榜    

查看: 170  |  回复: 0
当前主题已经存档。

俊铭

金虫 (正式写手)

[交流] 【求助放弃】翻译一篇文献的摘要和前言两段

取消求助翻译一篇文献的摘要和前言两段

要求:1 语序必须通顺
    2 专业一点,专业词汇可用英文单词代替
   3 可借助金山词霸,但大意还需一致。
   4 []内是参考文献,可以不翻,简写词也可不翻
  
Abstract
Advances in the theory of gradient liquid chromatography and their practical impacts are reviewed. Theoretical models describing retention in
reversed-phase, normal-phase and ion-exchange modes are compared. Main attention is focused on practically useful models described by twoor
three-parameter equations fitting the experimental data in the range of mobile phase composition utilized for sample migration during gradient
elution. The applications of theory for gradient method development, optimization and transfer are addressed. The origins and possibilities for
overcoming possible pitfalls are discussed, including the effects of the instrumental dwell volume, uptake of mobile phase components on the
column and size of the sample molecules. Special attention is focused on gradient separations of large molecules.

前言 Gradient elution versus isocratic elution
The profile of the gradient, i.e., the gradient steepness, the initial
and the final mobile phase composition, and in some cases the
shape of the gradient (curved or multi-segmented linear) affects
the retention of solutes in similar way as the elution strength in a
binary mobile phase under isocratic conditions [1,5–9]. This is
illustrated by the example of reversed-phase (RPC) separation of
lower n-alkylbenzenes under isocratic and gradient conditions.
Under isocratic conditions (Fig. 1), the retention times decrease
as the concentration of acetonitrile increases from 60 to 80% in
aqueous–organic mobile phases. Similar effect have increasing
gradient steepness (defined as the concentration change of the
strong solvent B in the gradient time from the start to the end of
the gradient, see Eq. (12)) and increasing initial concentration of
the solventB(acetonitrile) at the start of the gradient (see Fig. 2).
Both increasing gradient steepness and the initial concentration
of B decrease the retention times and result in narrower
peaks more regularly spaced in the chromatogram. This effect is
more apparent with gradients covering a broader concentration
range (cf. the separations in Fig. 2A, 60–100% acetonitrile, and
Fig. 2C, 80–100% acetonitrile. For comparison, see the isocratic
separation in 80% acetonitrile, Fig. 1C.
Many workers try to avoid gradient elution, as this technique
requires some additional time for column re-equilibration
between subsequent runs and because the selection of an optimumgradient
program and transfer of gradient methods between
the instruments and columns are less straightforward than the
development and transfer of isocratic separation methods, as
there are more parameters to be fixed. Further, higher-purity solvents
than in isocratic elution should be used in gradient-elution
to achieve high sensitivity in trace analysis [1]. Some detectors
are not compatible with gradient elution, such as most electrochemical
detectors or the refractometric detector. The latter one
is a universal detector, which gives a response for almost all sample
compounds, but also for the mobile phase components. The
only universal detector useful for gradient elution is the evaporative
light-scattering (ELS) detector, but it is approximately two orders of magnitude less sensitive than the UV detector [1,8].
As the ELS detector gives response to the stray light on solid
particles of analytes after evaporation of the solvent from the
nebulized column effluent, its use is restricted to volatile mobile
phases and nonvolatile analytes.

[ Last edited by telomerase on 2009-1-12 at 14:25 ]

» 猜你喜欢

已阅   关注TA 给TA发消息 送TA红花 TA的回帖
相关版块跳转 我要订阅楼主 俊铭 的主题更新
信息提示
请填处理意见