| 查看: 480 | 回复: 0 | |||
[交流]
Tzero技术是如何提升DSC性能:简化冷却试验的温度校准
|
| Because of thermal lag between DSC sample and temperature control sensors, most DSC’s require calibration at the heating rate to be used. Alternatively, calibration at multiple rates is required to correct for the heating rate dependence of thermal lag. Without this calibration there is considerable error in the temperature data in going from one heating rate to another. This problem is especially troublesome when performing a cooling experiment due to the thermal lag being in the opposite direct as that of heating doubling its effect. This problem is solved with a new DSC sensor employing TzeroTM technology. With the Q Series‰ DSC system an additional thermocouple on the DSC sensor allows complete calibration of the DSC cell and compensation for all thermal lags, including those caused by the encapsulated sample. The use of a liquid crystal sample is described to demonstrate that the calibration is independent of heating or cooling rate, even when using a non-ideal hermetic sample encapsulation. |
» 本帖附件资源列表
-
欢迎监督和反馈:小木虫仅提供交流平台,不对该内容负责。
本内容由用户自主发布,如果其内容涉及到知识产权问题,其责任在于用户本人,如对版权有异议,请联系邮箱:xiaomuchong@tal.com - 附件 1 : Tzero技术是如何提升DSC性能:简化冷却实验的温度校准.pdf
2017-03-01 11:05:51, 315.26 K
» 猜你喜欢
今年审到国自然15份,谈谈感受
已经有28人回复
投稿求助,期刊
已经有8人回复
光量子物理方向 博士招生 1人(2026.09)
已经有5人回复
河北省自然科学基金
已经有5人回复
献血感触
已经有13人回复
面上本子正文33页,违规吗?会被低分嘛?
已经有16人回复
国自然评分
已经有5人回复
国自然上会要求
已经有11人回复
提交了我也来说说感想
已经有10人回复
评审有感
已经有26人回复












回复此楼